Specifications
|
Toneohm 850A
|
|
RESISTANCE
|
Ranges |
High sensitivity ohm,
200mohm, 2 ohm, 200 ohm, 20k ohm |
Accuracy |
±4% in 200mohm, 2 ohm,
200 ohm, 20k ohm |
|
Approximately 40mohm
full scale on high sens. ohm range |
Probe voltage |
60mV maximum |
Probe protection |
Momentary contact up to
100V |
|
VOLTAGE
|
Ranges |
2mV, 20mV, 2V, 20V |
Accuracy |
±4% ±15µV |
Input resistance |
120 ohm in 2mV, 20mV |
|
1M ohm in 2V, 20V |
|
CURRENT TRACE
|
Sensitivity |
Detects current flow with
200 ohm connected to Drive Source |
|
DRIVE SOURCE
|
Output voltage |
0 to 550mV, adjustable |
|
POWER REQUIREMENTS
|
200-250V or 100-130V @
50/60Hz, 25VA |
|
ACCESSORIES (STANDARD)
|
Set of detachable probes |
|
Operator manual (English)
with User Guide in French, German and Italian |
|
APPROVALS
|
Conforms to applicable
European Directives and is CE marked Polar Instruments
Ltd is certified to ISO 9001, LRQA certificate number
932134. |
|
NATO STOCK NUMBER
|
Contact Polar Instruments |